Fluke 742A Resistance Standard Equivalency Guide
IET is pleased to offer a full line of equivalent resistors our SRL Series for those looking for the Fluke 742A Resistance Standards
IET Labs has been manufacturing standard resistors since 1976. Primary impedance standards: resistance, capacitance, and inductance are the core of our business.
Our products have evolved from the engineering behind GenRad and esi (Electro Scientific Industries) the two most respected names in impedance
Where IET Labs is not as large as a company or has as a worldwide recognized name as Fluke but in the world of impedance we are much respected and almost every NMI and primary lab in the world has a number of our products. The Fluke 742A is a very good resistor, from a great company but you should be confident that making resistors is what IET Labs do best and I can promise you that the SRL series should surpass your expectations.
Unlike the Fluke 742A which is only available at a couple, of values, IET SRL Resistors are available from any value from 1 mΩ - 1 TΩ.
You can see from the following chart, our SRL Resistors are equivalent or superior in every specification
IET LABS SRL SERIES VS FLUKE 742A | |||||||||
Datasheets can be found | |||||||||
Page 2 | Fluke 742a Manual | ||||||||
Page 3 and 4 | IET Labs SRL Datasheet | ||||||||
Model | Nominal Value | 6 Month Stability in ppm | 1 Year Stability in ppm | Calibration Uncertanty in ppm | Maximum Deviation from the 23 °C Value (18 to 28 °C) | +1 PPM Error Adder When Current Exceeds the Following | Max Voltage in V | Max Power W | Max Current A |
742-1 | 1 Ω | 5 | 8 | 1 | 3 | 200 mA | 0.5 | 0.25 | 0.5 |
SRL-1 | 1 Ω | 4 | 8 | 0.42 | 3 | 447 mA | 1 | 1 | 1 |
742-10 | 10 Ω | 5 | 8 | 1 | 3 | 100 mA | 1 | 0.1 | 0.1 |
SRL-10 | 10 Ω | 4 | 8 | 0.33 | 3 | 141 mA | 3 | 1 | 0.3 |
742-100 | 100 Ω | 5 | 8 | 1 | 3 | 5 mA | 2 | 0.4 | 0.2 |
SRL-100 | 100 Ω | 4 | 6 | 0.32 | 3 | 44 mA | 10 | 1 | 0.1 |
742-1K | 1 kΩ | 4 | 6 | 1.5 | 2 | 2 mA | 10 | 0.1 W | 0.01 |
SRL-1K-TC | 1 kΩ | 3 | 6 | 0.25 | 2 | 14 mA | 22 | 1 W | 0.22 |
742-10K | 10 kΩ | 2.5 | 4 | 1 | 1.5 | 600 µA | 30 | 0.09 | 3 mA |
SRL-10K | 10 kΩ | 2 | 4 | 0.19 | 1.5 | 4472 µA | 100 | 1 | 10 mA |
742-100K | 100 kΩ | 4 | 6 | 2.5 | 2 | 400 µA | 100 | 0.04 | 0.4 mA |
SRL-100K | 100 kΩ | 4 | 6 | 0.49 | 2 | 1414 µA | 316 | 1 | 3 mA |
742-1M | 1 MΩ | 6 | 8 | 5 | 2 | 100 µA | 100 | 0.01 W | 0.1 mA |
SRL-1M | 1 MΩ | 6 | 8 | 1 | 2 | 300 µA | 707 | 0.5 W | 0.7 mA |
742-10M | 10 MΩ | 6 | 9 | 10 | 3 | 20 µA | 200 | 0.04 W | 0.02 mA |
SRL-10M | 10 MΩ | 5 | 9 | 4.2 | 3 | 50 µA | 2236 | 0.5 W | 0.22 mA |
LOWER IS BETTER | LOWER IS BETTER | HIGHER IS BETTER |
Fluke Cross-Reference Guide | |
Fluke 720A Fluke Kelvin-Varley Divider | KVD-700 |
Fluke 742A-1 Fluke Resistance Standard | SRL-1 |
Fluke 742A-1.9 Fluke Resistance Standard | SRL-1.9 |
Fluke 742A-10 Fluke Resistance Standard | SRL-10 |
Fluke 742A-100 Fluke Resistance Standard | SRL-100 |
Fluke 742A-1K Fluke Resistance Standard | SRL-1K |
Fluke 742A-10K Fluke Resistance Standard | SRL-10K |
Fluke 742A-19K Fluke Resistance Standard | SRL-19K |
Fluke 742A-100K Fluke Resistance Standard | SRL-100K |
Fluke 742A-1M Fluke Resistance Standard | SRL-1M |
Fluke 742A-10M Fluke Resistance Standard | SRL-10M |
Fluke 742A-19M Fluke Resistance Standard | SRL-19M |